![Fenwal 4R5707 Operator'S Manual Download Page 360](http://html.mh-extra.com/html/fenwal/4r5707/4r5707_operators-manual_553093360.webp)
A
LYX
C
OMPONENT
C
OLLECTION
S
YSTEM
O
PERATOR
’
S
M
ANUAL
C
HAPTER
5
–
T
ROUBLESHOOTING
5-10
07-19-01-518 – January 2009
Typical Non-Recoverable Alarm with
Check Centrifuge Icon
Non-recoverable alarms that display a screen similar to the one
provided above indicate the instrument has detected a leak in the
centrifuge compartment.
WARNING: Do not attempt to reinfuse blood components
following fluid leaks or spills, when the instrument is
operating with unexpected noise or vibration, or when a
procedure is ended due to persistent redness in the plasma
tubing line coming from the centrifuge. In these instances,
the products must be discarded.
Under these circumstances, the operator should:
1. Press the STOP button.
2. Isolate the donor from the instrument by closing the clamp
on the donor line. The instrument will also engage the
donor safety clamp to isolate the donor when the alarm
sounds.
3. Touch the
end button
.
4. Follow the instructions/choices on the screens that follow in
order to end the procedure without fluid return. When a
leak is detected, no fluid return will be allowed.
5. Continue to follow the screen prompts to disconnect the
donor and remove the kit. Do NOT keep the product.
End Button
Summary of Contents for 4R5707
Page 2: ......