S
M
ANUAL
C
HAPTER
4A
–
A
LYX
2RBC-LR
C
OLLECTION
P
ROCEDURE
07-19-01-518 – January 2009
4A-33
Note:
The Hct/Hgb parameter can be defined to use either
hematocrit or hemoglobin. Hematocrit is used as the measurement
for this manual; however, hemoglobin may also be used.
13. Touch the
check button
after the donor’s hematocrit or
hemaglobin has been entered correctly.
After all donor information has been entered, the
red cell volume
button
will display the default procedure volume as specified by the
procedure nomogram (located in the Appendix) or center
requirements.
14. To adjust the target red cell volume, touch the
red cell volume
button
. Touch the button that displays the desired red cell
collection volume for the procedure.
Typical Select Target Red Cell Volume Overlay
15. Touch the
check button
to return to the Main Data Entry
Screen once the collection volume has been verified on the
screen overlay.
WARNING: Base the volume of the product(s) to be
collected on the maximum allowable target collection
volume(s) and on sound medical judgment, conforming to
applicable regulations and guidelines set by the appropriate
health regulating agencies.
Note:
Red cell target collection may have restricted volume
requirements depending on local guidelines.
Note:
The instrument will not allow entry of donor or product
inputs inconsistent with configured settings of the instrument’s
Administrative Settings. An audible tone and flashing data button
indicate entry of unacceptable data input.
Red Cell
Volume Button
Target Volume
Button
Check
Button
Check Button