Removing the Kit
The remove kit screen will be displayed on the touch screen.
Typical Procedure Information Screen with Unload
Kit (SUR)
1. Remove the red cell storage container from the scale and
label the product according to applicable standard operating
procedures.
Remove the rest of the kit from the instrument (Refer to Chapter
4A A
LYX
2RBC-LR Collection Procedure, “Removing the Kit”
for instructions regarding kit removal). Properly dispose of the
kit according to standard operating procedures.
2. Touch the
check button
to verify the kit has been removed
from the instrument.
Check Button
Unload Kit
Prompt