A
LYX
C
OMPONENT
C
OLLECTION
S
YSTEM
O
PERATOR
’
S
M
ANUAL
C
HAPTER
4B
–
A
LYX
RBC/P
LASMA
C
OLLECTION
P
ROCEDURE
07-19-01-518 – January 2009
4B-29
Note:
The Hct/Hgb parameter can be defined to use either
hematocrit or hemoglobin. Hematocrit is used as the
measurement for this manual; however, hemoglobin may also be
used.
13. Touch the
check button
after the donor’s hematocrit has been
entered correctly.
After all donor information has been entered, the
red cell volume
and
plasma volume buttons
will display the default procedure volume
as specified by the procedure nomogram (located in the
Appendix) or center requirements.
14. To adjust plasma volume, touch the
plasma volume button
and
select the desired plasma collection volume. Touch the
check
button
and verify the correct volume is displayed.
Typical Select Plasma Volume Overlay
15. Touch the
check button
to return to the Main Data Entry
Screen once the collection volume has been verified on the
screen overlay.
WARNING: Base the volume of the product(s) to be
collected on the maximum allowable target collection
volume(s) and on sound medical judgement, conforming to
applicable regulations and guidelines set by the appropriate
health regulating agencies.
Note:
The instrument will not allow entry of donor or product
inputs inconsistent with configured settings of the instrument’s
Administrative Settings. An audible tone and flashing data
button indicates entry of unacceptable data input.
Plasma Volume
Button
Check Button
Check Button