A
LYX
C
OMPONENT
C
OLLECTION
S
YSTEM
O
PERATOR
’
S
M
ANUAL
C
HAPTER
4D–
2RBC
S
INGLE
U
NIT
R
ECOVERY
P
ROCEDURE
4D-22
07-19-01-518 – January 2009
The operator ends the procedure by touching the STOP button
and selects the option to Process product (if available). The
remaining red cells from the in-process container and the
separation chamber will be processed and pumped to the red cell
container, if needed. Red cell preservation solution will
automatically be added to the red cell container. The mixture is
pumped through the leukoreduction filter and into one red cell
storage container.
Any collected red cells that exceed the maximum Single Unit
Recovery volume but are less than the minimum absolute red cell
volume for a 2RBC collection volume, will be pumped into the
in-process container or returned to the donor, if the ability to
return is an option.
1. Touch the STOP button to end the procedure.
Typical End Procedure Screen Overlay
An overlay screen similar to the one displayed above will be
shown with two possible options:
•
Return fluid to donor
•
Do not return fluid to donor
2. Close the clamp on one of the red cell storage containers.
3. Touch the
Do not return fluid to donor button
if no fluid return is
desired.
2RBC-LR Single
Unit Recovery
(SUR) Procedure –
Functionally Closed
Kit
Ending the
Procedure