A
LYX
C
OMPONENT
C
OLLECTION
S
YSTEM
O
PERATOR
’
S
M
ANUAL
C
HAPTER
4B
–
A
LYX
RBC/P
LASMA
C
OLLECTION
P
ROCEDURE
4B-18
07-19-01-518 – January 2009
19. Place the tubing elbow into the optical sensor, ensuring it is
flush in the slot, and press the end of the umbilicus into the
umbilicus holder.
Tubing Elbow and Umbilicus Properly Installed
20. Lift the kit tray to close the centrifuge door by pushing down
on the door until the handle latches.
WARNING: Do not operate the instrument if the
centrifuge door is open or damaged.
21. Remove the remaining containers and the donor line from
the kit tray and set them aside.
22. Open the cassette compartment door, remove the cassette
from the kit tray, and discard the tray.
WARNING: If the kit tubing or cassette is damaged,
discontinue the procedure.
23. Align the tabs on the top of the cassette with the cassette
guides on the instrument.
24. Press the cassette gently into place. Ensure the cassette fits
into all cassette guides on the instrument.
CAUTION:
Press on the rigid outer ridges of the cassette only.
Excessive pressure on the inner portion of the cassette may
damage the cassette.
Optical
Sensors
Umbilicus
Holder
Summary of Contents for 4R5707
Page 2: ......