The instrument will perform tests again to check that the
centrifuge door is open and the lock is functioning. If it detects
there is a problem with the door lock, a non-recoverable alarm,
(#7098) will occur and a screen similar to the one shown below
will be displayed. With the centrifuge door open, power cycle the
instrument OFF, wait at least 5 seconds, then turn it ON to see if
the alarm reoccurs. Contact the Hotline if this non-recoverable
alarm persists. Refer to the numerical listing of alarm codes for
additional information.
Typical Non-Recoverable Alarm Screen
Requiring Power Cycling
Other Non-Recoverable Alarms.
Other non-recoverable
alarms occur because the instrument has detected a serious
condition or failure that requires ending the procedure and
removing the kit from the instrument. The status zone displays
an alarm code number specific to the alarm and the icon on the
touch screen provides information about where the problems
may be found. Touch the
mute button
to temporarily silence the
alarm.
Note:
The
mute button
, when touched, will silence the alarm tone
for two minutes.
Mute Button