A
LYX COMPONENT COLLECTION
S
YSTEM
O
PERATOR
’
S
M
ANUAL
C
HAPTER
5
–
T
ROUBLESHOOTING
5-22
07-19-01-518 – January 2009
Code
Icons
Meaning
Resolution
4014
Non-Recoverable
Alarm Screen
The system has detected a software
error.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
5001 -
5006
Check Centrifuge
The system has detected an issue
with the centrifuge.
- Ensure separation chamber, umbilicus, and
elbow are properly installed.
- Ensure centrifuge door is properly closed.
- Resume the procedure.
- If problem persists, contact the Hotline.
7001
Non-Recoverable
Alarm Screen
Remove Kit
The system senses a change in
separation that may affect the
quality of the product. DO NOT
return fluids to the donor. Do not
keep the product.
- Press the STOP button to end the procedure.
- Disconnect the donor.
- Remove the kit.
- Discard the product.
7002
Non-Recoverable
Alarm Screen
Remove Kit
The system is having difficulty
removing air from the kit during
prime.
- Press the STOP button to end the procedure.
- Start a new procedure with a different kit.
7003 -
7004
Check Elbow
Check for Kink in
Line
The system is having difficulty
removing air from the kit during
prime. The elbow on the kit may
not be installed properly into the
centrifuge. There may be kinks in
the lines leading to the centrifuge.
- Open the centrifuge and check for proper
placement of the elbow.
- Close the centrifuge door.
- Check all tubing from the centrifuge to the
cassette for kinks or occlusions.
- Resume the procedure.
- If problem persists, press the STOP button to
end the procedure. Start a new procedure with a
different kit.
7005
Check for Kink in
Line
Check Cannulae
The system has detected an
unexpected event.
If during AC Prime, check that the cannula are
broken.
7006
Check Plasma Line
for Redness
The system has detected possible
lipemia, red cell spillover, or free
hemoglobin in the plasma line.
- Check the plasma line between the centrifuge
and the cassette for lipemia or redness.
- If only lipemia is found, resume the procedure.
- If redness is noted in the plasma line,
terminate the procedure without reinfusion
and discard the product.
- Contact The Hotline.
7007
Non-Recoverable
Alarm Screen
Check Plasma Line
for Redness
Remove Kit
The system has determined that it
cannot correct the red cell spillover,
or free hemoglobin is in the plasma
line.
- Terminate the procedure without
reinfusion and discard the product.
- Press the STOP button to end the procedure.
- Contact the Hotline.
7008
Check Cassette
The system has determined that the
cassette may not be installed
properly.
- Check that the cassette is installed properly
behind the cassette door.
- Check that the cassette door is completely
closed.
- Resume the procedure.