PERATING CONSIDERATIONS
07-19-01-518 – January 2009
3-79
Functionally Closed Kit
When the kit installation checks are complete, the touch screen
will display the kit timer and the
spike prompt
showing the
amount of time (minutes and seconds) remaining. Ensure there
is adequate time to complete all steps necessary before starting
the procedure.
Typical Spike Prompt and Kit Timer Screen
Functionally Closed Kit
CAUTION:
If the disposable kit is not primed, it should be
used within 8 hours after being loaded on the instrument.
However, once the cannulae have been broken on a closed kit,
or once kit-checks have started on a functionally closed kit, a
procedure must begin within 1 hour. After this time, the
instrument will alarm and prompt the operator to remove the
kit. The operator must press the STOP button, end the
procedure and remove the kit. A new procedure may then be
started.
CAUTION:
A disposable kit may only be reloaded as long as
the cannulae have not been broken, saline has not been spiked,
if applicable, and the sterility barrier(s) are intact. Otherwise if
the one-hour limit is exceeded, the kit must be discarded.