A
LYX
C
OMPONENT
C
OLLECTION
S
YSTEM
O
PERATOR
’
S
M
ANUAL
C
HAPTER
4D
-
2RBC
S
INGLE
U
NIT
R
ECOVERY
P
ROCEDURE
07-19-01-518 – January 2009
4D-17
To burp the red cell product, follow these steps:
1. Confirm that the clamp to the empty storage container is
closed.
2. Remove the paper strip from the red cell storage containers
segment tubing, if desired.
3. Leave the empty storage container on the hook. Remove the
filled storage container and hold port-side up.
4. Gently squeeze the filled container to expel any air into the
cassette.
5. Continue to squeeze the container until the air is removed
and the entire segment tubing is filled with red cells up to the
“X” just before reaching the “Y” junction.
6. Close the clamp on the line to the filled storage container.
Note
: If the air cannot be expelled when burping the red cell
container, check for kinks or occlusions in the line.
7. Touch the
check button
to verify the red cell product has been
burped.
Sealing the Kit
After burping the product, the Procedure Information Screen will
display the
seal kit prompt
.
Typical Procedure Information Screen with
Seal Kit Prompt (SUR)
Check Button
Seal Kit Prompt