Necessary optional devices will differ depending on the
vacuum pump to combine with.
-14-
4-3
Installation procedures
※
1 Diaphragm type vacuum pumps that can be connected are
as follows:
Models NVP-1000
・
2000
・
2100
Models EVP-1000
・
1100
・
1200
Models DTC-22
・
41, Model DTU-20, Model MD1C, Model
N920
(You cannot use any other vacuum pump types.)
※
2 Power for NVC-3000 is supplied from the model PBX.
※
3 Never attempt to touch the surface of CV-11 or CV-12
solenoid valve for NVC control since it may be hot during
operation.
※
4 Do not install the PBX with its panel surface (connector
surface) facing above to avoid foreign objects or water from
entering.
4-3-1
When a diaphragm type vacuum pump is used
NVC-PBX connecting
cord
(
Optional
)
Use optional items below to combine the model NVC-3000 with a diaphragm
type vacuum pump (constant rotation)
・
Diaphragm pump control box PBX
・
NVC control solenoid valve
Model CV-11 (For 1 to 5 litter condensation units)
Model CV-12 (For 5 to 10 litter condensation units)
・
NVC-PBX connecting cord
NP0.5M
(length:0.5m)
NP2M
(length:2m)
Model PBX
(
Optional)
Model CV-11
Model CV-12
(Optional)
※
3
(1) Connect the model NVC-3000 and the model PBX as shown in
the right figure with a connecting cord.
(2) Connect the connector for the CV solenoid valve for NVC
control to the model PBX.
(3) Connect the Teflon tube included with the NVC control
solenoid valve to the connecting port of the vacuum line of
NVC-3000.
(
Refer to the instruction manual of the CV-11 & CV-12)
(4) Insert the connecting cord to the outlet of the model PBX
and insert the power plug of the diaphragm type vacuum
pump into the opposite socket.
※
1
(5) Insert the power cord included with the model PBX into
the Power inlet with a fuse of the model PBX.
Teflon tubes CV-11 &
12 included
Vacuum line
connecting port
Power inlet with a fuse
Power cord
Pump connecting
cord
Outlet
Connector for
NVC control
solenoid
connector
Connect to the
diaphragm type
vacuum pump
suction nozzle
Vacuum
hose
Connect to the
decompression
container
(6) Insert the vacuum hose into the nozzle of the NVC control
solenoid valve to the end.
Carefully check where to connect the vacuum hose beforehand.
Insert both NVC-3000
and PBX so that their
arrows are at the top
PBX panel surface
※
4