-40-
■
System Parameter 1 Setting screen
(5) Manual Cleaning Time
(
Manual Cleaning Time
)
Setting range
:
1
~
10min(unit:1)
Initial value
:
3 min
You can set a manual cleaning time.
(4) Setting of a Cleaning Start Time
(
Cleaning Start Timer
)
Setting range
:
0
~
60sec (unit:1)
Initial value
:
30 sec
You can set a wait time from completion of auto leak after
operation before auto cleaning starts.
(6) Control of the coolant circulation pump
(
Circulate Pump Control
)
Setting range
:
Manual, Auto
Initial setting
:
Manual
Control for the circulation pump of the coolant circulation unit when
it is connected with the NVC communication cable.
Manual setting: Does not control the circulation pump.
Auto setting: Circulation pump is turned ON at the start of operation
and the pump is turned OFF when operation is stopped.
See P.43 “Controlling related units connected via NVC
communication” for details.
(7) Auto stop control of the bath temperature control(Bath Auto Stop
Control
)
Setting range
:
Manual, Auto
Initial setting
:
Manual
You can set auto stop of temperature control for the bath
interlocked with NVC or the evaporator when the water bath or
the oil bath is connected with the NVC communication cable.
Manual setting:Does not perform auto stop of temperature control.
Temperature control is stopped when “(12)
System Alarm” occurred.
Auto setting: Stops temperature control of the bath interlocked
with the operation stop of NVC. When the
evaporator is connected with the NVC
communication cable, bath temperature control is
stopped about 2 seconds after turning the
evaporator power OFF.
* Bath temperature control can be started from the bath only.
Leak
Menu
Run
Stop
Power
End
SYSTEM PARAMETER 1
Leak
Menu
Run
Stop
Power
Evap Jack Up/Down
Auto Leak
On
Auto Cleaning 3 min
Cleaning Start Timer 30 sec
Manual Cleaning Time 3 min
Circulate Pump Control Manual
Bath Auto Stop Control Manual