NVC communication
cable
Vacuum control
communication
cable
Use a communication cable (optional) and a branching box (optional) to combine
the NVC-3000 and a related unit such as an evaporator with a communication
terminal.
Product name
Connection
Additional functions
Vacuum control communication
cable
VC-2M
(
length:2m
)
VC-4M
(
length:4m
)
※
Note that you cannot use
this cable when you use the
inverter pump.
Connect the vacuum
controller model NVC-
3000 and the model NVC-
3000 vacuum controller.
When two vacuum controllers are controlled with one
diaphragm vacuum pump, you can operate these units by
monitoring their operation and by preventing from interfering
with operations of each other.
NVC communication cable
COM-0.5M
(
length:0.5m
)
COM-2M
(
length:2m
)
Connect the model NVC-
3000 vacuum controller to
a condensation unit-related
unit with a communication
terminal.
Evaporator
:
N-1300, N-1210B, N-2110
You can start, stop rotation of, set and display the number of
rotations from NVC-3000.
Water/oil bath:SB-1200,OSB-2200
You can stop temperature control interlocked with operation
and display a temperature from NVC-3000.
(
Starting and setting of temperature control is possible from
the bath only.
)
Coolant circulation unit
:
CA-1116A, CA-1330, CCA-1112A
You can control the circulation pump interlocked with
operation and set and display temperatures from NVC-3000.
Branching box F-BOX for NVC
communication
Branches the NVC
communication cable
The NVC communication is branching-connected when you
connect two or more condensation related units to NVC-3000.
The specifications are the same for 4 connectors.
-17-
4-5
Connecting the communication cable (optional)
NVC communication
branching box F-BOX
Communication terminals for
the evaporator, the water/oil
baths, the coolant circulation
unit
NVC
communication
cable
Comm. connector
NVC-NVC
connector
Communication terminals for
the evaporator, the water/oil
baths, the coolant circulation
unit
Communication terminals for
the evaporator, the water/oil
baths, the coolant circulation
unit
(When branching is
unnecessary)
Do not switch off the inter-connected
units with the communication cables
If the inter-connected unit is switched off
during the operation ,it may cause the interruption
of the communication between the Vacuum Controller
and the inter-connected units
Caution