-24-
■
Control mode select screen
■
Graph screen
■
Setting screen
Leak
Menu
Run
Stop
Power
MANUAL
00:04:46
Esc
Set Press
hPa
Leak
Menu
Run
Stop
Power
200
1000
0
15
min
Run
50
℃
200
Leak
Menu
Run
Stop
Power
MANUAL
Standby
00:00:00
Esc
hPa
Leak
Menu
Run
Stop
Power
980
Set Press
200
hPa
Δ P Auto
hPa
Auto Stop Off
Leak
Menu
Run
Stop
Power
MANUAL
Standby
00:00:00
Esc
1013
hPa
Set Press
hPa
200
Leak
Menu
Run
Stop
Power
Set Press
Time
Pressure
Auto Stop
5-3 Operation in the MANUAL (fixed-
value control) mode
After operation is started, fixed-value control will start at
the fixed pressure (Set Press).
You can stop operation of the unit automatically
according to the operation time by setting an operation
stop condition (Auto Stop).
You can stop the unit automatically based on the
temperature increase margin from the vapor temperature
by using a vapor temperature sensor (optional).
(1) Fixed pressure (Set Press) setting
Setting range
:
VAC
、
1
~
1013hPa(unit:1)
Initial value
:
200 hPa
VAC will be controlled but continuously decreased.
you can change settings in any mode of standby,
decompression, or fixed-value control and in any of [Control
mode select screen], [Graph screen], or [Setting screen].
●
In the Control mode select screen or the Graph screen
Turn the dial key to start setting.
This [Set Press] color will change from light blue to yellow
and you can change the value using the dial.
Turn the dial key to change the value to the pressure you
want and then press the key to select it as the fixed-
pressure.
(Once a value is determined, its color will return from
yellow to light blue.)
* When five seconds have passed without
determining a value, it will return to the one before
change.
●
In the Setting screen
Press the [Display] key on the Control mode select screen
to change to the Setting screen.
Turn the dial key to move the cursor to [Set Press] and
press it to start setting.
(The indication color will change from light blue to
yellow.)
Turn the dial key to change the value to the pressure you
want and then press the key to select it as the fixed-
pressure.
(Once a value is determined, its color will return from
yellow to light blue.)
*
You can quickly change the values by turning the
dial key faster.