(1)
Set pressure (Press
)
Setting range
:
VAC, 1
~
1013hPa(unit:1), Atmo
Initial value :VAC
VAC Decompression is made continuously.
Atmo The leak valve is opened and decompression is
released.
* You can quickly change the values by turning
the dial key faster.
(2)
Set time (Time)
Setting range
:
0
~
999min(unit:1)
Initial value
:
0
* You can quickly change the values by turning
the dial key faster.
* If you set Time to [0] for slope control, pressure
will change to the set pressure at the highest
rate.
(3)
Hysteresis (control width) (ΔP
)
Setting range
:
Auto, 1
~
20hPa(unit:1)
Initial value
:
Auto
In the Auto setting, automatic control is performed with
2% of the set pressure set as ΔP.
* No setting is necessary for the inverter pump
.
-35-
■
Control mode select screen
■
Program setting screen
Leak
Menu
Run
Stop
Power
Mode
STEP PROGRAM
Standby
00:00:00
Display
1013
hPa
Leak
Menu
Run
Stop
Power
P No
1
REPEAT
0 / 1
(4) Control condition select (Condition
)
Setting conditions
:
Step, Slope
Initial value
:
Step
Step
Performs fixed value control at the set pressure for
the set duration.
Slope Controls the slope so that it follows the line that
connects the slope start pressure (set press of the
previous step) and the set pressure based on the set
time. When Slope is set for the first step after
operation has started, the pressure measured at that
time will be set as the slope start pressure.
Leak
Menu
Run
Stop
Power
End
00:00:00
Display
Leak
Menu
Run
Stop
Power
STEP PROGRAM
Standby
P No
1
hPa
1013
Press(hPa)
Time (min)
ΔP (hPa)
Condition
Priority
Step No 1 2 3 4 5
VAC
0
Auto
Step
Press
VAC
0
Auto
Step
Press
VAC
0
Auto
Step
Press
VAC
0
Auto
Step
Press
VAC
0
Auto
Step
Press
Leak
Menu
Run
Stop
Power
00:00:00
Esc
Leak
Menu
Run
Stop
Power
STEP PROGRAM
Standby
P No
1
hPa
1013
Press(hPa)
Time (min)
ΔP (hPa)
Condition
Priority
Step No 1 2 3 4 5
VAC
0
Auto
Step
Press
VAC
0
Auto
Step
Press
VAC
0
Auto
Step
Press
VAC
0
Auto
Step
Press
VAC
0
Auto
Step
Press
You can set a program on the Program setting screen.
Turn the dial key to move the cursor to the item you are
going to make setting for.
*
Turning the dial key clockwise will move the
cursor in the following order:
StepNo
→ Press → Time → ΔP →
Condition
→
Priority
→
next step
・・・
Press the dial key to start setting.
(The indication color will change from light blue to
yellow.)
Turn the dial key to change the value to the value you want
and press to determine.