-32-
(4) User solvents
You can set solvent names and settings as user solvents in
No.56
~
65 of the database according to your objectives.
The initial settings are as follows.
Change them to your specific objectives.
Fixed pressure (Set Press) : 200 [hPa]
Slope starting pressure (Slope Point): 400 [hPa]
Slope time (Slope Time) : 5 [min]
Hysteresis (Δ P) : Auto
Operation stop condition (Auto Stop) : Off
Stop temperature (Stop Temp) : 3.5 [
℃
]
Solvent name (Name) : User-01
~
User-10
Settings and solvent names can be changed in the setting
screen.
Follow the procedures below to display the setting screen.
Select a user solvent to change in the database screen.
(“User-01” in the upper right example)
Press the [End] key to change to the solvent register screen.
Move the cursor to the solvent to change, press the dial
key to change to “SELECT DATA” for solvents.
Then press the dial key again to change to change to the
setting screen.
Items whose settings can be changes are as follows.
●
Fixed value pressure (Set Press)
Setting range : VAC , 1
~
1013 hPa (Scale:1)
●
Slope start pressure (Slope Point)
Setting range : 1
~
1013 hPa (Scale:1)
●
Slope time (Slope Time)
Setting range : 1
~
999 min (Scale:1)
●
Hysteresis (ΔP)
Setting range : Auto , 1
~
20 hPa (Scale:1)
*No settings are necessary for the inverter pump.
●
Operation stopping condition (Auto Stop)
Setting range: Off , 1
~
999 min (Scale:1)
Temp (When a vapor temperature
sensor is connected)
●
Stop temperature (Stop Temp)
It is valid when the operation stopping condition (Auto
Stop) is set to Temp.
Setting range : 0.1
~
15.0
℃
(Scale:0.1)
●
Solvent name (Name)
You can set a solvent name up to 18 alphanumeric
characters, space and symbols ( ,.-+*/() ).
See P.33 for setting procedures.
■
Solvent registration screen
Leak
Menu
Run
Stop
Power
SlctAllClr
PROGRAM
Standby
00:00:00
End
hPa
Leak
Menu
Run
Stop
Power
1013
DATA BASE 11
Acetone
56 User-01
57 User-02
58 User-03
59 User-04
60 User-05
61 User-06
Leak
Menu
Run
Stop
Power
DataBase
PROGRAM
Standby
00:00:00
End
hPa
Leak
Menu
Run
Stop
Power
1013
SELECT DATA 11
User-01
2 Acetone
56 User-01
7 Benzene
13 Chloroform
17 1,2-Dichloroethane
21 Dichloromethane
26 Diethyl ether
Leak
Menu
Run
Stop
Power
End
PROGRAM
Standby
00:00:00
Back
hPa
Leak
Menu
Run
Stop
Power
1013
User-01
Set Press
200
hPa
Slope Point 400 hPa
Slope Time 5 min
ΔP Auto hPa
Auto Stop Off
Stop Temp 3.5
℃
Name User-01
Vapor 27.6
℃
■
Database screen
■
Setting screen
* When a vapor temperature sensor is
connected
Leak
Menu
Run
Stop
Power
DataBase
PROGRAM
Standby
00:00:00
End
hPa
Leak
Menu
Run
Stop
Power
1013
SELECT DATA 11
2 Acetone
56 User-01
7 Benzene
13 Chloroform
17 1,2-Dichloroethane
21 Dichloromethane
26 Diethyl ether
Acetone