No
Indications and names
Contents and status
How to release
06
Data error
EEPROM DATA
This error occurs when there is an
error in the data in the non-volatile
memory when power is turned on.
The portion with that error will be
reset to the factory shipping settings.
Use the [Clr] key to clear errors.
The display will return to the
ordinary screen.
The buzzer, if it is sounding, will
stop.
07
Memory write error
EEPROM WRITE
This error is triggered when a write
error of the user data region to the
non-volatile memory occurs.
(Settings including pressure, program
and calibration data)
Use the [Clr] key to clear errors.
The display will return to the
ordinary screen.
The buzzer, if it is sounding, will
stop.
09
NVC mutual
communication error
NVC-NVC
This error is triggered when the
mutual communication between
vacuum control assemblies is
impossible for more than five seconds
during operation (Run, Leak, or
Cleaning).
This error occurs when connection of
the inverter pump is detected during
mutual connection with NVC.
This error occurs when a unit other
than the model NVC-3000 (such as
the model NVC-2300) is connected.
The unit will stop at the operation end
status.
(Only when the vacuum control
assembly mutual communication
function is used)
Use the [Clr] key to clear errors.
The display will return to the
ordinary screen.
The buzzer, if it is sounding, will
stop.
10
Inverter pump alarm
Inverter Pump
This error is triggered when a error
occurs to the inverter pump.
The unit will stop at the operation end
status.
(Only when the inverter pump is
used)
Use the [Clr] key to clear errors.
The display will return to the
ordinary screen.
The buzzer, if it is sounding, will
stop.
11
Control valve wrong
connection alarm
Control Valve
This error occurs when CV-1 or CV-2 is
connected to the control valve.
This error status stop when operation is
finished.
Use the [Clr] key to clear errors.
The display will return to the
ordinary screen.
Only the buzzer if it is sounding will
stop.
12
System Alarm
System Alarm
This error occurs when an alarm
generated on a related unit connected by
communication is received.
Operation stops without releasing
vacuum.
(Only when the related unit is connected
by communication.)
Use the [Clr] key to clear errors.
The display will return to the
ordinary screen.
The buzzer, if it is sounding, will
stop.
ALARM
(06) EEPROM DATA
Clr
ALARM
(07) EEPROM WRITE
Clr
ALARM
(09) NVC-NVC
Clr
ALARM
(10) Inverter Pump
Clr
ALARM
(12) System Alarm
Clr
ALARM
(11) Control Valve
Clr