-39-
■
Control mode select screen
■
System Parameter 1 Setting screen
■
System Parameter 2 Setting screen
Leak
Menu
Run
Stop
Power
End
SYSTEM PARAMETER 2
Leak
Menu
Run
Stop
Power
Key Buzzer
On
Dial Buzzer On
ALARM Buzzer On
Running Buzzer On
Back Light 5
MAIN/SUB Screen MAIN
Leak
Menu
Run
Stop
Power
Mode
MANUAL
Standby
00:00:00
Display
1013
hPa
Set Press
hPa
200
Leak
Menu
Run
Stop
Power
Leak
Menu
Run
Stop
Power
End
SYSTEM PARAMETER 3
Pressure Offset 0 980 hPa
Vapor Temp Offset 0.0 27.8
℃
Pressure Unit hPa
Temperature Unit
℃
User DATA BASE Reset
RESET
All DATA BASE Reset
Memory Initialize INITIAL
Leak
Menu
Run
Stop
Power
■
System Parameter 3 Setting screen
System Parameter 1 Setting screen [SYSTEM PARAMETER 1]
(1) Operation of the evaporator jack (Evap Jack Up/Down)
This screen is displayed while N-1210B or N-2110 is
connected.
Press the dial key to highlight this item in yellow, and click
the key once to perform jack up or jack down for a short time.
Turn the key clockwise to jack up and anticlockwise to jack
down.
To finish, press the dial key to return the color to light blue.
Leak
Menu
Run
Stop
Power
End
SYSTEM PARAMETER 1
Leak
Menu
Run
Stop
Power
Evap Jack Up/Down
Auto Leak
On
Auto Cleaning 3 min
Cleaning Start Timer 30 sec
Manual Cleaning Time 3 min
Circulate Pump Control Manual
Bath Auto Stop Control Manual
5-7
Setting parameters
You can make settings of various operations or reset data you
have changed on the system parameter setting screen.
You can move to the system parameter setting screen by
pressing [Menu] key.
There are three types of the system parameter setting screens.
Switch the indication with the [Menu] key and press the
[End] key to return to the Control mode select screen.
* The System parameter 3 setting screen will not
appear during operation.
(3) Auto Cleaning (Auto Cleaning
)
Setting range
:
Off, 1
~
10min(unit:1)
Initial value
:
3 min
You can set an Auto cleaning time after Auto Leak has
completed.
When Off is selected, Auto cleaning will not be carried out
after Auto Leak.
* Step program operation is controlled according to the
setting on the Program operation setting screen.
【
Caution
】
In principle, do not set Auto cleaning Off because it
purges liquefied steam from the control solenoid valve
or the diaphragm vacuum pump.
(Set a time.)
(2) Automatic Leak (Auto Leak)
Setting range: On, Off
Initial value
:
On
You can set the release of decompression at the end of operation.
When you select On, Auto Leak will be carried out at the end of
operation.
When you select Off, Auto Leak and Auto Cleaning will not be
carried out at the end of operation.
* Step program operation is controlled according to the
setting on the Program operation setting screen.
【
Caution
】
Note that auto cleaning will not operate while Auto Leak is
set at Off.
In principle, set Auto Leak to On.