-22-
(3) Pressing the [Display] key displays the graph screen and
pressing it again displays the setting screen.
* See the following pages for how to operation each control
mode.
* You can return to the control mode select screen by
pressing the [End] key on the setting screen.
(4) Items that you can change are shown in light blue in each
screen.
Once you start adjusting an item by pressing the dial, the item
will change to yellow.
The item will return to light blue when it is determined by
pressing the dial.
(5) After having set operating conditions such as pressure,
temperature or time for the control mode you want to use, use
the Run/Stop key to start operation.
■
Setting screen
Leak
Menu
Run
Stop
Power
Mode
MANUAL
Standby
00:00:00
Display
1013
hPa
Set Press
hPa
200
Leak
Menu
Run
Stop
Power
Leak
Menu
Run
Stop
Power
Mode
MANUAL
Standby
00:00:00
End
hPa
Leak
Menu
Run
Stop
Power
1013
Set Press
200
hPa
Δ P Auto
hPa
Auto Stop Off
■
Graph screen
Leak
Menu
Run
Stop
Power
Mode
MANUAL
00:00:00
Display
Set Press
hPa
200
Leak
Menu
Run
Stop
Power
1013
1000
0
15
min
Standby
■
Control mode select screen
(6) Communication system control screen
Changing the setting in the MAIN/SUB Screen of the system
parameter 2 screen (See P.41) will change the screen from the
control mode select screen to the communication system
control screen. Setting to
MAIN+SUB will display both the
control mode select screen and the communication system
control screen.
Settings and measured values are also displayed for the related
units that are connected and communicated with the NVC
communication cable.
You can change the set vacuum level, the set rotation of the
evaporator and the set temperature of the coolant circulation
unit in this screen.
*
Settings changed from NVC will remain valid as long as
communication with the NVC continues.
The settings will return to those before change in about 1
minute after communication is shut off.
Leak
Menu
Run
Stop
Power
Mode
MANUAL
Standby
00:00:00
Display
Leak
Menu
Run
Stop
Power
Vacuum
200
980
hPa
Rotation 180 0
rpm
Circulator 5 0.0
℃
Bath
40
40.0
℃
Vapor
27.6
℃
■
Communication system control screen