22
5.2 Connecting the product
1. Unscrew the protective cap from the probe coupling.
2. Screw the cryoprobe to the cryosurgery unit via the connect-
ing adapter (Art. No. 20416-034).
5.3 Checking
function and seal
Perform the check in a well-lit location.
1. Immerse the cryotip to a depth of at least 5 cm in sterile
warm water at a temperature of approx. 20 °C (alternatively:
isotonic NaCl solution).
2. If small bubbles develop during immersion: Diffuse the bub-
bles by briskly moving the cryotip back and forth in the wa-
ter.
3. Activate the freeze function of the cryosurgical unit for ap-
prox. 5 seconds.
A clearly visible, homogenous ball of ice must form at the cryotip
when freezing capacity is sufficient. Gas bubbles must not es-
cape under any circumstances.
Never use a product that is not leak-tight or a product with in-
adequate freezing capacity. There is a risk that a gas embolism
could develop if an application element is used that is not leak-
tight.
5.4 Using the product
Note:
To prevent the cryotip from freezing to the inside of the
endoscope, the metal cryotip should be fully extended outside of
the working channel of the endoscope throughout use. The first
black ring must be visible.
OBJ_DOKU-181800-009.fm Seite 22 Freitag, 14. Juni 2019 6:50 06
Summary of Contents for 20416-032
Page 2: ...OBJ_DOKU 181800 009 fm Seite 2 Freitag 14 Juni 2019 6 50 06...
Page 87: ...EL 20416 032 20416 037 OBJ_DOKU 181800 009 fm Seite 87 Freitag 14 Juni 2019 6 50 06...
Page 91: ...91 4 1 2 3 4 5 5 5 1 1 1 2 3 4 5 OBJ_DOKU 181800 009 fm Seite 91 Freitag 14 Juni 2019 6 50 06...
Page 93: ...93 2 3 4 5 1 2 3 4 OBJ_DOKU 181800 009 fm Seite 93 Freitag 14 Juni 2019 6 50 06...
Page 100: ...100 Erbe 7 8 x X CE OBJ_DOKU 181800 009 fm Seite 100 Freitag 14 Juni 2019 6 50 06...
Page 184: ...184 OBJ_DOKU 181800 009 fm Seite 184 Freitag 14 Juni 2019 6 50 06...
Page 199: ...RU 20416 032 20416 037 OBJ_DOKU 181800 009 fm Seite 199 Freitag 14 Juni 2019 6 50 06...
Page 205: ...205 1 2 3 4 5 1 2 3 4 OBJ_DOKU 181800 009 fm Seite 205 Freitag 14 Juni 2019 6 50 06...
Page 212: ...212 Erbe 7 8 x X CE OBJ_DOKU 181800 009 fm Seite 212 Freitag 14 Juni 2019 6 50 06...
Page 227: ...JA 20416 032 20416 037 OBJ_DOKU 181800 009 fm Seite 227 Freitag 14 Juni 2019 6 50 06...
Page 230: ...230 40 60 bar 60 bar OBJ_DOKU 181800 009 fm Seite 230 Freitag 14 Juni 2019 6 50 06...
Page 233: ...233 3 4 5 1 2 3 4 5 OBJ_DOKU 181800 009 fm Seite 233 Freitag 14 Juni 2019 6 50 06...
Page 240: ...240 x X CE OBJ_DOKU 181800 009 fm Seite 240 Freitag 14 Juni 2019 6 50 06...
Page 246: ...246 5 4 1 2 3 4 5 1 2 3 4 5 OBJ_DOKU 181800 009 fm Seite 246 Freitag 14 Juni 2019 6 50 06...
Page 252: ...252 8 x CE X CE OBJ_DOKU 181800 009 fm Seite 252 Freitag 14 Juni 2019 6 50 06...
Page 253: ...KO 20416 032 20416 037 OBJ_DOKU 181800 009 fm Seite 253 Freitag 14 Juni 2019 6 50 06...
Page 256: ...256 O Erbe Erbe OBJ_DOKU 181800 009 fm Seite 256 Freitag 14 Juni 2019 6 50 06...
Page 265: ...265 Erbe 7 8 batch x X CE OBJ_DOKU 181800 009 fm Seite 265 Freitag 14 Juni 2019 6 50 06...
Page 266: ...266 OBJ_DOKU 181800 009 fm Seite 266 Freitag 14 Juni 2019 6 50 06...
Page 267: ...OBJ_DOKU 181800 009 fm Seite 267 Freitag 14 Juni 2019 6 50 06...