iTHERM TrustSens TM371, TM372
Mounting
Hauser
11
U
≥
3°
≥ 3°
1
2
3
4
A0031007
3
Mounting possibilities in the process
1, 2 Perpendicular to flow direction, installed at a min. angle of 3° to ensure self-draining
3
On elbows
4
Inclined installation in pipes with a small nominal diameter
U
Immersion length
In the case of pipes with a small nominal diameter, it is advisable for the tip of the
thermometer to project well into the process so that it extends past the pipe axis.
Installation at an angle (4) could be another solution. When determining the immersion
length or installation depth all the parameters of the thermometer and of the medium to
be measured must be taken into account (e.g. flow velocity, process pressure).
Maximum torque
1
2
3
4
5
M
M
M
A0035951
Protection tube version TT411, 6 mm (0.24 in) (1)
TT411, 6 mm (0.24 in) and
Necktube TE411 (2)
TT411,
9 mm (0.35 in) (3)
TT411, 12.7 mm (¹⁄₂ in) (4)
TT411, 12.7 mm (¹⁄₂ in) and
Necktube TE411 (5)
Torque M
3 to 5 Nm (2.2 to 3.7 lbf ft)
10 Nm (7.4 lbf ft)
3 to 5 Nm (2.2 to 3.7 lbf ft)
When connecting the device with the protection tube: only turn the hexagonal
spanner flat on the bottom of the housing.