Device functionality checked on demand
• Verification of the correct functioning of the measuring device within specifications.
• The verification result provides information about the condition of the device: "Passed" or
"Failed".
• The results are documented in a verification report.
• The automatically generated report supports the obligation to demonstrate compliance
with internal and external regulations, laws and standards.
• Verification is possible without interrupting the process.
Advantages
• No onsite presence is required to use the function.
• The DTM
triggers verification in the device and interprets the results. No specific
knowledge is required on the part of the user.
• The verification report can be used to prove quality measures to a third party.
•
Heartbeat Verification
can replace other maintenance tasks (e.g. periodic check) or
extend the test intervals.
Detailed description
Heartbeat Monitoring
Function
Calibration information is logged in addition to the verification parameters. 350
calibration points are saved in the device (FIFO memory).
Advantages
• Early detection of changes (trends) to ensure plant availability and product quality.
• Use of information for the proactive planning of measures (e.g. maintenance).
Detailed description
8.6.2
Heartbeat Diagnostics
Device diagnostic messages, along with remedial measures, are displayed in the operating
tool (FieldCare/DeviceCare).
For information on using the diagnostic messages, see the "Diagnostics and
troubleshooting" section. → 35
Diagnostic message in the operating tool
1. Navigate to the "
Diagnostics
" menu.
The diagnostic event, along with the event text, is displayed in the
Actual
diagnostics
parameter.
1)
DTM: Device Type Manager; controls device operation via DeviceCare, FieldCare, PACTware or a DTM-based control system.