iTHERM TrustSens TM371, TM372
Commissioning
Hauser
33
Test object
Verification criterion
Sensor module
Sensor
Checks whether the sensor works as per the specifications.
Reference temperature
Checks whether the reference sensor works as per the specifications.
Sensor drift warning limit
exceeded
Checks whether the configured warning limits are exceeded.
Sensor drift alarm limit exceeded Checks whether the configured alarm limits are exceeded.
Sensor information
Number of self-calibrations
Displays all the self-calibrations executed up to now. This value cannot be
reset.
Deviation
Displays the deviation of the measured value from the reference
temperature.
Adjustment of the measurement
Displays the adjustment of the calibration deviation.
Monitoring parameters
Device temperature min:
Displays the minimum electronics temperature measured in the past
(peakhold indicator).
Device temperature max:
Displays the maximum electronics temperature measured in the past
(peakhold indicator).
Sensor min value:
Displays the minimum temperature measured in the past at the sensor
input (peakhold indicator).
Sensor max. value:
Displays the maximum temperature measured in the past at the sensor
input (peakhold indicator).
Summary of results
Overall
results
Displays the overall result of the verification. The verification report can be saved in PDF and XML
format. To save the report, click the
Save results as PDF
button or the
Save results as XML
button.
If the verification fails, try again or contact the Service Organization.
8.6.4
Heartbeat Monitoring
Calibration information is logged in addition to the verification parameters.
HART variable
Output
Unit
PV
Temperature
°C/°F
SV
Device temperature
°C/°F
TV
Calibration counter
-
QV
Calibration deviation
°C/°F
Monitoring information can be read out and analyzed as described below:
A higher-level controller is configured in such a way that calibration deviations and the
calibration counter are saved when the calibration counter changes. This type of function
is supported by Hauser's Advanced Data Manager Memograph M RSG45, for
example. The following table provides a sample overview of the monitoring analysis using
the Field Data Manager software MS20: