iTHERM TrustSens TM371, TM372
Operating menu and parameter description
Hauser
81
Navigation
Calibration → Limits → Upper warning value
Description
Enter the upper warning limit for the self calibration deviation.
User entry
+0.5 to +1.0 · 10
20
°C
Factory setting
+0.5 °C
Additional information
Use this function to define the upper warning limit. If the self-calibration deviation
exceeds the defined limit, the device will transmit the defined status signal and shows the
defined diagnostic behavior via the LED.
(
Factory setting
= Warning - LED red flashes).
Lower alarm value
Navigation
Calibration → Limits → Lower alarm value
Description
Enter the lower alarm limit for the self-calibration deviation.
User entry
–1.0 · 10
20
to –0.8 °C
Factory setting
–0.8 °C
Additional information
Use this function to define the lower alarm limit. If the self-calibration deviation exceeds
the defined limit, the device will transmit the defined status signal and show the defined
diagnostic behavior via the LED (diagnostic event 143).
(
Factory setting
= Warning - LED red flashes).
Upper alarm value
Navigation
Calibration → Limits → Upper alarm value
Description
Enter the upper alarm limit for the self-calibration deviation.
User entry
+0.8 to +1.0 · 10
20
°C
Factory setting
+0.8 °C
Additional information
Use this function to define the upper alarm limit. If the self-calibration deviation exceeds
the defined limit, the device will transmit the defined status signal and shows the defined
diagnostic behavior via the LED.
(
Factory setting
= Warning - LED red flashes).