Commissioning
iTHERM TrustSens TM371, TM372
26
Hauser
Navigation
"Calibration" menu → Limits → Lower alarm value
"Calibration" menu → Limits → Upper alarm value
1. In the
Lower alarm value
input window, enter the lower limit for the self-calibration
deviation and press ENTER to confirm.
2. In the
Upper alarm value
input window, enter the upper warning limit for the self-
calibration deviation and press ENTER to confirm.
Please consider the maximum limited value range to be entered: ±0.35 °C (±0.67 °F).
8.4
Creating a calibration report
The "calibration report" wizard guides you systematically through the process for creating a
calibration report for a pre-selected calibration point.
Navigation
"Calibration" menu → Calibration report
At least one stored self-calibration point must be in the device to start the online
wizard.
Configuration and creation of a calibration report
1.
1.
2.
A0033678-EN
To read the calibration point data from the device, enter the calibration point index.
Index 1 reads the latest calibration point.
2. Press READ DATA to confirm.
An overview of device information and calibration point data appears. See table
below for detailed information.
Summary of Contents for iTHERM TrustSens TM371