iTHERM TrustSens TM371, TM372
Technical Data
Hauser
53
T / °C (°F)
t
123 °C
(253.4 °F)
> 25 s
< 116 °C (240.8 °F)
-16.5 K/min
(-29.7 °F/min)
t1
t2
1
t3
t3
116 °C
(240.8 °F)
-0.5 K/min
(-0.9 °F/min)
2
A0032839
10
Needed process temperature profile for self-calibration
1
Process temperature 123 °C (253.4 °F)
2
Allowed self-calibration range
Calibration monitoring
Available in conjunction with Advanced Data Manager Memograph M (RSG45). → 47
Application package:
• Up to 20 devices can be monitored via the HART interface
• Self-calibration data displayed on screen or via the Web server
• Generation of a calibration history
• Creation of a calibration protocol as an RTF file directly at the RSG45
• Evaluation, analysis and further processing of the calibration data using "Field Data
Manager" (FDM) analysis software
Insulation resistance
Insulation resistance ≥ 100 MΩ at ambient temperature.
Insulation resistance between the terminals and the outer jacket is measured with a
minimum voltage of 100 V DC.
13.5 Environment
Ambient temperature
range
Ambient temperature T
a
–40 to +60 °C (–40 to +140 °F)
Maximum electronics
temperature T
–40 to +85 °C (–40 to +185 °F)
Storage temperature range
T = –40 to +85 °C (–40 to +185 °F)
Climate class
As per IEC 60654-1, Class Dx
Summary of Contents for iTHERM TrustSens TM371