A. Introduction

B. WARNING & CAUTIONS

C. Distance to spot size

    This infrared thermometer is used for measuring the tem-
perature of  the object's surface, which is applicable for va-
rious hot, hazardous or hard-toreach objects without contact 
safely and quickly. 
    This unit consist of Optics, Temperature Sensor Signal 
amplifier, Processing circuit and LCD Display. The Optics 
collected the infrared energy emitted by object and focus onto 
the Sensor. Then the sensor translates the energy into an 
electricity signal. This signal will be turned out to be digital 
shown on the LCD after the signal amplifier and processing 
circuit. 

1. Warning:      
    To avoid the potential situation may cause harm or dam-
age to people, please pay attention to the following items:
1) . Do not point laser directly at eye or indirectly off reflec-
      tive surfaces.
2) . The unit cannot measure through transparent surfaces 
      such as glass or plastic. It will measure the surface tem-
      perature of these materials instead.
3) . Steam, dust, smoke, or other particles can prevent acc-
      urate measurement by obstructing by the units optics.
 
2. Cautions:
    Infrared thermometer should be protected for the 
following: 
1). EMF (electro-magnetic fields) from arc welders, 
      induction heaters.
2). Thermal shock (cause by large or abrupt ambient 
      temperature changes allow 30 minutes for unit to 
      stabilize before use). 
3). Do not leave the unit on or near objects of high 
      temperature.

1. When take measurement, pay attention to the Distance 
    to Spot Size. As the Distance (D) from the target surface 
    increases, the spot size (S) of the area measured by the
    unit becomes larger.
    The Distance to Spot size of the unit is 12:1(Figure 1).

MODEL:WT320

Version

WT320-EN-0

Infrared Thermometer

Instruction Manual

 

D. EMISSIVITY

    Emissivity: Most organic materials and painted or 
oxidized surfaces have an emissivity of 0.95(preset in 
the unit). Inaccurate readings will result from measuring 
shiny or polished metal surfaces. To compensate for this, 
adjust the units emissivity reading or cover the surface to 
be measured with masking tape or flat black paint. 
Measure the tape or painted surface when the tape or 
painted reach the same temperature as the material 
underneath.

2. Field of view: 
    Make sure the target is larger than the unit's spot size. 
The smaller the target the closer measure distance. 
When accuracy is critical, make sure the target is at least 
twice as large as the spot size. 

E. Operation

1.  Operating the unit

Figure 2

1). Open the battery door and insert the battery.
2). Pull the trigger to turn on the unit;
3).  Aim at the target surface and pull the trigger, then temp-
     erature will be shown on the LCD. 
    This unit is equ ipped with a laser, which is only used for 

 

      aiming.  
    

Trigger

Finger indents

Figure 2

25mm@
300mm

75mm@
900mm

125mm@
1500mm

1"@
12"

3"@
36"

 5"@
 60"

D:S=12:1