Deltabar FMD72
Description of device parameters
101
Min. meas. press. (266)
Navigation
Diagnosis
®
Sensor LP
®
Minimum Pressure (266)
Write permission
No write permissions. Parameter is read only.
Description
Displays the lowest pressure value measured (peakhold indicator). You can reset this indicator by
means of the "Reset peakhold" parameter.
COUNTER P < Pmin (267)
Navigation
Diagnosis
®
Sensor LP
®
Counter P < Pmin (267)
Write permission
No write permissions. Parameter is read only.
Description
Displays the negative pressure counter for the respective sensor module. The counter is incremented
each time error 841 occurs. You can reset this value using the "Reset Peakhold (161)" parameter.
Max. meas. press. (268)
Navigation
Diagnosis
®
Sensor LP
®
Maximum Pressure (268)
Write permission
No write permissions. Parameter is read only.
Description
Displays the highest pressure value measured (peakhold indicator). You can reset this peak indicator
via the "Reset peakhold (161)" parameter.
COUNTER P > Pmax (269)
Navigation
Diagnosis
®
Sensor LP
®
Counter P > Pmax (269)
Write permission
No write permissions. Parameter is read only.
Description
Displays the overpressure counter for the respective sensor module. The limit value is: upper sensor
module nominal value + 10% of upper sensor module nominal value. You can reset this value using
the "Reset Peakhold (161)" parameter.
MIN. MEAS. TEMP. (270)
Navigation
Diagnosis
®
Sensor LP
®
Min. Meas.Temp. (270)
Summary of Contents for Deltabar FMD72
Page 2: ......
Page 107: ......
Page 108: ...www endress com worldwide 71191239 71191239 BA01044P 00 EN 02 12 71191239 CCS COSIMA...