Integrating transmitter using HART
®
protocol
Deltabar FMD72
38
10.2
Device variables and measured values
The following measured values are assigned to the individual device variables:
Device variable code
Measured value
0
Measured differential pressure
1
Corrected pressure
2
Measured pressure HP
3
Sensor press. HP
4
Sensor temperature HP
5
Measured pressure LP
6
Sensor press. LP
7
Sensor temperature LP
8
Level before linearization
9
Tank content
10
Process density
11
Electronics temperature
12
HART input value
The device variables can be queried from a HART
®
master using HART
®
command 9 or 33.