Description of device parameters
Deltabar FMD72
80
Meas.Diff.Press. (020)
Navigation
Setup
®
Meas.Diff.Press. (020)
Diagnosis
®
Measured Values
®
Meas.Diff.Press. (020)
Write permission
No write permissions. Parameter is read only.
Description
Displays the measured differential pressure after sensor trim, position adjustment and damping.
P
HP
LP
Pressure
simulation
HP
Differential
pr
pressure inversion
Position
correction
(Calib.offset)
Diff.
pressure
simulation
Damping
Corrected
press.
Pressure
after
damping
Meas.
diff. pressure
Meas.
pressure HP
Meas.
pressure LP
Pressure
simulation
LP
Sensor
Trimm
HP
Sensor
calibration
HP
Sensor pressure HP
Sensor pressure LP
Sensor
Trimm
LP
Sensor
calibration
LP
A0017360-EN
Operator Code (021)
Navigation
Setup
®
Advanced Setup
®
Operator Code (021)
Write permission
Operators/Service engineers/Expert
Description
Use this function to enter a code to lock or unlock operation.
User entry
• To lock: Enter a number
¹
the release code (value range: 1 to 9999).
• To unlock: Enter the release code.
Note
The release code is "0" in the order configuration. Another release code can be defined in the "Code
Definition" parameter. If the user has forgotten the release code, the release code can be visible by
entering the number "5864".
Factory setting
0
Device tag (022)