Description of device parameters
Deltabar FMD72
76
Options
• No value
• Differential pressure
• Pressure HP
• Pressure LP
• Sensor Temp. HP
• Sensor Temp. LP
• Level before linearization
• Current
• Main measured value (%)
The options depend on the measuring mode chosen.
Factory setting
No value
Format 1st Value (004)
Navigation
Display/operat.
®
Format 1st Value (004)
Write permission
Operators/Service engineers/Expert
Description
Specify the number of places after the decimal point for the value displayed in the main line.
Options
• Auto
• x
• x.x
• x.xx
• x.xxx
• x.xxxx
• x.xxxxx
Factory setting
Auto
Operating mode (005/182)
Navigation
Setup
®
Measuring Mode (005/182)
Write permission
Operators/Service engineers/Expert
Description
Select the measuring mode. The operating menu is structured differently depending on the
measuring mode selected.
Options
• Pressure
• Level
Note
If the measuring mode is changed, no conversion takes place. If necessary, the device has to be
recalibrated after the measuring mode has been changed.
Factory setting
Level or according to order specifications