Deltabar FMD72
Commissioning
47
Description
11
Use the "Set URV (167)" parameter to set the level
value for the upper current value (20 mA) (3 m (9.8
ft)).
Menu path: Setup
®
Extended Setup
®
Current
Output
®
Set URV
12
If the process uses a medium other than that on
which the calibration was based, the new density
must be specified in the "Process Density (035)"
parameter.
Menu path: Setup
®
Extended Setup
®
Level
®
Process Density.
13
Result:
The measuring range is set for 0 to 3 m (9.8 ft).
For this level mode, the measured variables %, level, volume and mass are available, see "Unit
before lin. (025)" (
11.7.4 "In pressure" level selection Calibration without reference
pressure (dry calibration)
Example:
In this example, the volume in a tank should be measured in liters. The maximum volume of
1000 liters (264 gal) corresponds to a pressure of 450 mbar (6.53 psi). The minimum volume of 0
liters corresponds to a pressure of 50 mbar (0.73 psi) since the device is mounted below the start of
the level measuring range.
Prerequisite:
• The measured variable is in direct proportion to the pressure.
• This is a theoretical calibration i.e. the pressure and volume values for the lower and upper
calibration point must be known.