Deltabar FMD72
Operation with an operating menu
35
Write permission
Operators/Service engineers/Expert
Description
Use this function to enter a code to lock or unlock operation.
User entry
• To lock: Enter a number
¹
the release code (value range: 1 to 9999).
• To unlock: Enter the release code.
Factory setting
0
Note
The release code is "0" in the order configuration. Another release code can be defined in the "Code
Definition" parameter. If the user has forgotten the release code, the release code can be visible by
entering the number "5864".
The release code is defined in the "Code Definition" parameter.
Code Definition (023)
Navigation
Setup
®
Advanced Setup
®
Code Definition
Read permission
Operators/Service engineers/Expert
Write permission
Operators/Service engineers/Expert
Description
Use this function to enter a release code with which the device can be unlocked.
User entry
A number from 0 to 9999
Factory setting
0
9.9
Resetting to factory settings (reset)
By entering a certain code, you can completely, or partially, reset the entries for the parameters to
the factory settings
. Enter the code via the "Reset" parameter (menu path: "Diagnosis"
®
"Reset").
There are various reset codes for the device. The following table illustrates which parameters are
reset by the particular reset codes. Operation must be unlocked to reset parameters (see "Locking/
unlocking operation" section (
1)
. The factory setting for the individual parameters is specified in the parameter description
Summary of Contents for Deltabar FMD72
Page 2: ......
Page 107: ......
Page 108: ...www endress com worldwide 71191239 71191239 BA01044P 00 EN 02 12 71191239 CCS COSIMA...