Deltabar FMD72
Description of device parameters
81
Navigation
Setup
®
Advanced Setup
®
Device tag (022)
Write permission
Operators/Service engineers/Expert
Description
Enter the device tag, e.g. TAG number (max. 32 alphanumeric characters).
Factory setting
No entry or according to order specifications
Code Definition (023)
Navigation
Setup
®
Advanced Setup
®
Code Definition (023)
Write permission
Operators/Service engineers/Expert
Description
Use this function to enter a release code with which the device can be unlocked.
Options
A number from 0 to 9999
Factory setting
0
Level Selection (024)
Navigation
Setup
®
Advanced Setup
®
Level
®
Level Selection (024)
Write permission
Operators/Service engineers/Expert
Description
Select the method for calculating the level
Options
• In pressure
If this option is selected, specify two pressure/level value pairs. The level value is displayed directly
in the unit that you select via the "Unit Before Lin." parameter.
• In height
If this option is selected, specify two height/level value pairs. From the measured pressure, the
device first calculates the height using the density. This information is then used to calculate the
level in the "Unit Before Lin." selected using the two value pairs specified.
Factory setting
In pressure
Unit before lin. (025)
Navigation
Setup
®
Advanced Setup
®
Level
®
Unit Before Lin. (025)
Write permission
Operators/Service engineers/Expert
Description
Select the unit for the measured value display for the level before linearization.