Deltabar FMD72
Commissioning
43
11.5.3 Measuring mode selection
Measuring mode (005)
Navigation
Setup
®
Measuring mode
Write permission
Operators/Service engineers/Expert
Description
Select the measuring mode.
The operating menu is structured differently depending on the measuring mode selected.
Options
• Pressure
• Level
Factory setting
Level
Note
If the measuring mode is changed, no conversion takes place. If necessary, the device has to be
recalibrated after the measuring mode has been changed.
11.5.4 Defining the high-pressure side
High Press. Side (183)
Navigation
Setup
®
High Press. Side
Write permission
Operators/Service engineers/Expert
Description
Define which sensor module corresponds to the high-pressure side.
Options
• Sensor module HP
• Sensor module LP
Factory setting
Sensor module HP
11.5.5 Press. Eng. Unit selection
Press. Eng. Unit (125)
Navigation
Setup
®
Press. Eng. Unit
Write permission
Operators/Service engineers/Expert
Description
Select the pressure engineering unit. If a new pressure engineering unit is selected, all pressure-
specific parameters are converted and displayed with the new unit.
Summary of Contents for Deltabar FMD72