3.14.2
Installation steps
Follow these steps to install the HTS remote threaded seal on an existing process flange:
Procedure
1. Remove the protective cover from the diaphragm of the remote seal. Use extreme
caution during installation to ensure the diaphragm is not damaged.
2. Apply the thread sealant to the male thread.
3. Thread the remote seal onto the process connection. The remote seal is designed
with a small hole to facilitate installation and applying torque. The applied torque
should comply with ANSI B1.20.1 for NPT connections or applicable torque
requirements for pipe connections.
3.15
SCW hygienic Tri-Clover Tri Clamp seals
Figure 3-18: SCW Tri Clamp Seal
2
½
-, 3-, and 4-in. size
A
B
2-in. size
1
½
-in. size
A
B
A
B
A. Connection to transmitter
B. Diaphragm
Installation
Reference Manual
May 2020
00809-0100-4002
48
Rosemount DP Level
Summary of Contents for Rosemount 1199
Page 4: ...4 ...
Page 8: ...Introduction Reference Manual May 2020 00809 0100 4002 8 Rosemount DP Level ...
Page 22: ...Remote Seal Systems Reference Manual May 2020 00809 0100 4002 22 Rosemount DP Level ...
Page 74: ...Installation Reference Manual May 2020 00809 0100 4002 74 Rosemount DP Level ...
Page 92: ...Fill Fluids Reference Manual May 2020 00809 0100 4002 92 Rosemount DP Level ...