Potential cause
Cold temperature
Recommended actions
Fill fluid viscosity is temperature dependent. Less viscous fill fluid enhances time
response. Heat traced capillaries can be added as an option to maintain constant
temperatures to fill fluid.
6.2.3
Drifting
Potential cause
Drifting
Recommended actions
If pressure measurement is changing, refer to
information. Running Instrument Toolkit will calculate the expected performance for
the seal system. Refer to
Instrument Toolkit: seal ordering and application process
for
more information.
6.2.4
Output reads negative pressure
Potential cause
Mounting effect
Recommended actions
The output will read negative as the remote fill fluid is applying pressure to the low
side. See
6.2.5
No response to change in pressure
Potential cause
Damaged diaphragm
Recommended actions
Remove seal and inspect diaphragm.
6.2.6
Not responding to change in pressure
Potential cause
Multidrop mode
Recommended actions
See if the transmitter is in a multidrop mode.The multidrop mode locks the output at
4 mA.