3.15.3
Installation steps
Follow these steps to install the SCW hygienic Tri Clamp seal:
Procedure
1. Remove the protective cover from the diaphragm of the remote seal. Use extreme
caution during installation to ensure the diaphragm is not damaged.
2. Place the appropriate gasket for the application on the remote seal and align the
gasket so that it is not contacting the diaphragm as this will induce errors.
WARNING
Failure to properly install the gasket may cause process leaks, which can result in
death or serious injury.
3. Ensure that the SCW hygienic Tri Clamp seal is flush with the process connection.
4. Install clamp and tighten to manufacturer's specifications
Note
EHEDG requirement on T-stub installations: the length (L) divided by the diameter
(D) of the tee is to be less than 1 (i.e. L /D < 1).
D
L
Installation
Reference Manual
May 2020
00809-0100-4002
50
Rosemount DP Level