3.17
STW hygienic thin wall tank spud seal
Figure 3-24: STW Hygienic Thin Wall Tank Spud Seal
A
B
A. Connection to transmitter
B. Diaphragm
3.17.1
Parts required for installation
The parts required to install the STW hygienic thin wall tank spud seal will be defined here.
Prior to installation, you will need the following:
•
O-ring (Emerson-supplied)
•
Clamp (Emerson-supplied)
Cutting and welding equipment required if installing a new tank spud.
Verify the O-ring material is appropriate for the application.
3.17.2
Installation steps
Follow these steps to install the STW hygienic thin wall tank spud seal:
Procedure
1. Remove the plastic cover from the diaphragm of the remote seal.
2. Weld the tank spud onto the tank per plant procedures, see
.
WARNING
Ensure the tank spud is not assembled to the pressure transmitter and/or the
remote seal prior to welding.
3. Remove the protective cover from the diaphragm of the remote seal. Use extreme
caution during installation to ensure the diaphragm is not damaged.
4. Place the Emerson-supplied O-ring onto the groove. Failure to properly install the
gasket may cause process leaks, which can result in death or serious injury.
5. Insert the STW hygienic thin wall tank spud seal into the tank spud.
6. Attach the Emerson-supplied clamp and hand-tighten the connection.
Reference Manual
Installation
00809-0100-4002
May 2020
Rosemount DP Level
55
Summary of Contents for Rosemount 1199
Page 4: ...4 ...
Page 8: ...Introduction Reference Manual May 2020 00809 0100 4002 8 Rosemount DP Level ...
Page 22: ...Remote Seal Systems Reference Manual May 2020 00809 0100 4002 22 Rosemount DP Level ...
Page 74: ...Installation Reference Manual May 2020 00809 0100 4002 74 Rosemount DP Level ...
Page 92: ...Fill Fluids Reference Manual May 2020 00809 0100 4002 92 Rosemount DP Level ...