3.20
SVS VARIVENT
®
compatible hygienic
connection seal
Figure 3-27: SVS VARIVENT Compatible Connection Seal
A
B
A. Connection to transmitter
B. Diaphragm
3.20.1
Parts required for installation
The parts required to install the SVS VARIVENT seal will be defined here. Prior to
installation, you will need the following:
•
Clamp (end-user-supplied)
•
Gasket (end-user-supplied)
•
Wrench (optional, depending on clamp)
Verify the gasket material is appropriate for the application. Ensure to use an EHEDG
approved gasket if EHEDG conformance is needed.
3.20.2
Installation steps
Follow these steps to install the SVS VARIVENT seal:
Procedure
1. Remove the protective cover from the diaphragm of the remote seal. Use extreme
caution during installation to ensure the diaphragm is not damaged.
2. Place the appropriate gasket for the application on the remote seal.
WARNING
Failure to properly install the gasket may cause process leaks, which can result in
death or serious injury.
3. Ensure that the SVS VARIVENT seal is flush with the process connection.
4. Install clamp and tighten to manufacturer's specifications.
Reference Manual
Installation
00809-0100-4002
May 2020
Rosemount DP Level
59