Configure device options and preferences
Table 6-1
Status alarms and
Status Alarm Severity
continued
Alarm
code
Status message
Default
severity
Notes
Configurable?
A105
Slug flow
Informational
Yes
A106
Burst mode enabled
Informational
Can be set to either
Informational
or
Ignore
, but cannot be set to
Fault
.
Yes
A107
Power reset occurred
Informational
Normal transmitter behavior;
occurs after every power cycle.
Yes
A108
Event 1 triggered
Informational
Applies only to basic events.
Yes
A109
Event 2 triggered
Informational
Applies only to basic events.
Yes
A110
Frequency output saturated
Informational
Can be set to either
Informational
or
Ignore
, but cannot be set to
Fault
.
Yes
A111
Frequency output fixed
Informational
Can be set to either
Informational
or
Ignore
, but cannot be set to
Fault
.
Yes
A113
Secondary mA output saturated
Informational
Can be set to either
Informational
or
Ignore
, but cannot be set to
Fault
.
Yes
A114
Secondary mA output fixed
Informational
Can be set to either
Informational
or
Ignore
, but cannot be set to
Fault
.
Yes
A115
External input error
Informational
Yes
A116
API temperature outside standard
range
Informational
Applies only to transmitters with
the petroleum measurement
application.
Yes
A117
API density out of limits
Informational
Applies only to transmitters with
the petroleum measurement
application.
Yes
A118
Discrete output 1 fixed
Informational
Can be set to either
Informational
or
Ignore
, but cannot be set to
Fault
.
Yes
A120
Concentration measurement:
unable to fix curve data
Informational
Applies only to transmitters with
the concentration measurement
application.
No
A121
Concentration measurement:
extrapolation alarm
Informational
Applies only to transmitters with
the concentration measurement
application.
Yes
Configuration and Use Manual