Configure device options and preferences
Option
Description
Enabled
(default)
Operator can access the maintenance section
of the off-line menu. This access is required
for configuration and calibration, but is not
required to view alarms.
Disabled
Operator cannot access the maintenance
section of the off-line menu.
2.
To control operator access to the alarm menu, enable or disable
Alarm Menu
.
Option
Description
Enabled
(default)
Operator can access the alarm menu. This
access is required to view and acknowledge
alarms, but is not required for configuration,
or calibration.
Disabled
Operator cannot access the alarm menu.
Note
The status LED on the transmitter face shows whether or not alarms are active, but does not show
specific alarms.
3.
Enable or disable
Off-Line Password
as desired.
Option
Description
Enabled
Operator is prompted for the off-line password
at entry to the maintenance section of the
off-line menu.
Disabled
(default)
No password is required for entry to the
maintenance section of the off-line menu.
4.
Enable or disable
Alarm Password
as desired.
Option
Description
Enabled
Operator is prompted for the off-line password
at entry to the alarm menu.
Disabled
(default)
No password is required for entry to the alarm
menu.
If both
Off-Line Password
and
Alarm Password
are enabled, the operator is prompted for the off-line
password at the top of the off-line menu, and is not prompted thereafter.
5.
(Optional) Set
Off-Line Password
to the desired value.
The default value is 1234. The range is 0000 to 9999.
106
Micro Motion 9739 MVD Transmitters