Configure device options and preferences
Restrictions
• For some alarms,
Status Alarm Severity
is not configurable.
• For some alarms,
Status Alarm Severity
can be set only to two of the three options.
Procedure
1.
Select a status alarm.
2.
Set
Status Alarm Severity
as desired.
Option
Description
Fault
Actions when alarm is detected:
• Alarm posted to Alert List
• Outputs go to configured fault action (after
Fault Timeout
has expired, if applicable)
• Digital communications go to configured
fault action (after
Fault Timeout
has expired,
if applicable)
• Status LED changes to red or yellow
(depending on alarm severity)
Actions when alarm clears:
• Outputs return to normal behavior
• Digital communications returns to normal
behavior
• Status LED changes behavior (returns to
green, may or may not flash)
Informational
Actions when alarm is detected:
• Alarm posted to Alert List
• Status LED changes to red or yellow
(depending on alarm severity)
Actions when alarm clears:
• Status LED changes behavior (returns to
green, may or may not flash)
Ignore
No action
Status alarms and options for
Status Alarm Severity
Each status alarm has a default
Status Alarm Severity
. Some status alarms can be configured for other
severity levels.
Configuration and Use Manual
109