Transmitter operation
e.
Select
again to confirm.
f.
Scroll
to
EXIT
.
8.6
Reset mass and volume totalizers
Display
See Section 8.6.1 .
ProLink II
ProLink→Totalizer Control→Reset Mass Total
ProLink→Totalizer Control→Reset Volume Total
ProLink→Totalizer Control→Reset Gas Volume Total
ProLink→Totalizer Control→Reset
Field Communicator
Service Tools→Variables→Totalizer Control→Mass→Mass Total
Service Tools→Variables→Totalizer Control→Gas Standard Volume→Volume Total
Service Tools→Variables→Totalizer Control→Gas Standard Volume→GSV Total
Service Tools→Variables→Totalizer Control→All Totalizers→Reset All Totals
8.6.1
Reset mass and volume totalizers using the display
Prerequisites
To reset the totalizers using the display, this feature must be enabled and the appropriate process
variable (
Mass Total
,
Volume Total
, or
Gas Volume Total
) must be configured as a display variable.
Procedure
• To reset the mass totalizer using the display:
a.
Scroll
until the mass totalizer value appears.
b.
Select
.
c.
Scroll
until
RESET
appears beneath the current totalizer value.
d.
Select
.
e.
Select
again to confirm.
f.
Scroll
to
EXIT
.
g.
Select
.
• To reset the volume (liquid or gas) totalizers using the display:
a.
Scroll
until the volume totalizer value appears.
b.
Select
.
c.
Scroll
until
RESET
appears beneath the current totalizer value.
d.
Select
.
e.
Select
again to confirm.
f.
Scroll
to
EXIT
.
g.
Select
.
162
Micro Motion 9739 MVD Transmitters