Configure device options and preferences
Procedure
Enable or disable
Backlight
as desired.
The default setting is
Enabled
.
6.2
Enable or disable operator actions from the display
Display
OFF-LINE MAINT→OFF-LINE CONFG→DISPLAY
ProLink II
ProLink→Configuration→Display→Display Options
Field Communicator
Not available
You can control whether or not the operator will be able to perform specific actions from the transmitter
display.
You can:
• Enable or disable
Totalizer Start/Stop
• Enable or disable
Totalizer Reset
• Enable or disable
Acknowledge All Alarms
6.2.1
Enable or disable
Totalizer Start/Stop
from the display
Display
OFF-LINE MAINT→OFF-LINE CONFG→DISPLAY→TOTALS STOP
ProLink II
ProLink→Configuration→Display→Display Options→Display Start/Stop Totalizers
Field Communicator
Not available
You can control whether or not the operator will be able to start and stop totalizers and inventories
from the display.
Restrictions
• You cannot start and stop totalizers individually from the display. When you use the display to start
or stop totalizers, all totalizers are started or stopped together.
• You cannot start or stop inventories separately from totalizers. When a totalizer is started or
stopped, the associated inventory is also started or stopped.
• If the petroleum measurement application is installed on your computer, the operator must enter
the off-line password to perform this function, even if the off-line password is not enabled.
Procedure
1.
Enable or disable
Totalizer Start/Stop
as desired.
Configuration and Use Manual
103