Quick start with ProLink II
Figure 3-7
ProLink II connection to a multidrop network
B
C
A
A
HART interface (to PC)
B
Devices on the network
C
Master device
6.
Start ProLink II.
7.
Choose
Connection→Connect to Device
.
8.
If necessary, set the connection parameters to match the HART/RS-485 parameters configured in
your transmitter.
To minimize configuration requirements, the 9739 MVD transmitter uses an auto-detection scheme
when responding to a connection request. The transmitter will accept all connection requests
within the limits described in Table 3-1 .
Table 3-1
Auto-detection limits
Parameter
Option
Protocol
HART, Modbus ASCII, Modbus RTU
Address
Responds to:
•
Service port address (111)
•
Configured HART address (default = 0)
•
Configured Modbus address (default = 1)
Baud rate
Standard rates between 1200 and 38,400
Stop bits
0, 1
Parity
Even, odd, none
9.
Set the
COM Port
value to the PC COM port that you are using for this connection.
36
Micro Motion 9739 MVD Transmitters