Configure device options and preferences
Procedure
Set
Update Period
as desired.
The default value is 200 milliseconds. The range is 100 milliseconds to 10,000 milliseconds
(10 seconds).
6.1.5
Enable or disable automatic scrolling through the display
variables
Display
OFF-LINE MAINT→OFF-LINE CONFG→DISPLAY→AUTO SCRLL
ProLink II
ProLink→Configuration→Display→Display Options→Display Auto Scroll
Field Communicator
Not available
You can configure the display to show a single display variable indefinitely (until the operator activates
Scroll
), or to scroll through the configured display variables and display each one for a user-defined
number of seconds.
Procedure
1.
Enable or disable
Auto Scroll
as desired.
Option
Description
Enabled
The display will automatically scroll through the
list of display variables, showing each display
variable for the number of seconds specified
by
Scroll Rate
. The operator can move to the
next display variable by activating
Scroll
.
Disabled
(default)
The display will show
Display Variable 1
and will
not scroll automatically. The operator can
move to the next display variable by activating
Scroll
.
2.
If you enabled
Auto Scroll
, set
Scroll Rate
as desired.
The default value is 10 seconds.
Tip
You may need to apply the
Auto Scroll
setting before you can access
Scroll Rate
.
6.1.6
Enable or disable the display backlight
Display
OFF-LINE MAINT→OFF-LINE CONFG→DISPLAY→BKLT
ProLink II
ProLink→Configuration→Display→Display Options→Display Backlight On/Off
Field Communicator
Not available
You can enable or disable the backlight on the display’s LCD panel.
102