Quick start with the Field Communicator
Figure 4-1
Field Communicator connection to HART clips
C
B
A
A
Field Communicator
B
250–600 Ω resistance
C
Transmitter
2.
To connect to a point in the local HART loop, attach the leads from the Field Communicator to any
point in the loop and add resistance as necessary. See Figure 4-2 .
The Field Communicator must be connected across a resistance of 250–600 Ω.
Figure 4-2
Field Communicator connection to local HART loop
A
B
C
A
Field Communicator
B
250–600 Ω resistance
C
Transmitter terminals
Configuration and Use Manual