SEP 2005
MODEL 1000
DESCRIPTION
1-14
1.6.3
Peak Detection
For normal area or peak height concentration evaluation, the determination of a peak's start, peak
point, and end is automatic. The manual determination of start and end points is used only for area
calculations in the Forced Integration mode. Automatic determination of peak onset or start is
initiated whenever Integrate Inhibit is turned off. Analysis is started in a region of signal quiescence
and stability, such that the signal level and activity can be considered as baseline values. It is
important that this be the case because the assumption is made by the GC Controller software.
Having initiated a peak search by turning Inhibit off, the GC Controller performs a point by point
examination of the signal slope. This is achieved by using a digital slope detection filter which is
a combination low pass filter and differentiator. The output of this detector is constantly compared
to a system constant entered by the operator called Slope Sensitivity. A default value of 8 is assumed
if no entry is made. Lower values make peak onset detection more sensitive, and higher values make
detection less sensitive. Higher values (20 to 100) would be appropriate for noisy signals, e.g. high
amplifier gain.
Peak termination is determined by the same application of this detector to the signal, but in the
reverse sense. Onset is defined where the detector output exceeds the baseline constant, but
termination is defined subsequently where the detector output is less than the same constant.
Sequences of fused peaks are also automatically handled. This is done by testing each termination
point to see if the region immediately following it satisfies the criteria of a baseline. A baseline
region must have a slope detector value less than the magnitude of the baseline constant for a number
of sequential points. When a baseline region is found, this terminates a sequence of peaks.
A zero reference line for peak height and area determination is established by extending a line from
the point of the onset of the peak sequence to the point of the termination. The values of these two
points are found by averaging the four integrated points just prior to the onset point and just after the
termination points, respectively. The zero reference line will, in general, be non-horizontal, and thus
compensates for any linear drift in the system from the time the peak sequence starts until it ends.
Summary of Contents for Daniel Danalyzer 1000
Page 2: ......
Page 18: ...SEP 2005 MODEL 1000 DESCRIPTION 1 6 Figure 1 1 GC System Functional Block Diagram ...
Page 38: ...SEP 2005 MODEL 1000 EQUIPMENT DESCRIPTION 2 6 Figure 2 1 Analyzer Components ...
Page 146: ...SEP 2005 MODEL 1000 INSTALLATION AND SETUP 3 92 Figure 3 37 Analyzer Components ...
Page 188: ...SEP 2005 MODEL 1000 MAINTENANCE 5 20 Figure 5 2 Analyzer Components ...
Page 222: ...SEP 2005 MODEL 1000 MAINTENANCE 5 54 This page intentionally left blank ...
Page 228: ...SEP 2005 MODEL 1000 LIST OF APPENDICES A ii This page intentionally left blank ...
Page 260: ...SEP 2005 MODEL 1000 SETTING SOLENOID PURGE FLOWS E 4 This page intentionally left blank ...
Page 282: ...SEP 2005 MODEL 1000 ANALYZER DRAWINGS Addendum 1 2 This page intentionally left blank ...
Page 283: ......
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ......
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......
Page 299: ......
Page 300: ......
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......
Page 311: ......
Page 312: ......
Page 313: ......
Page 314: ......
Page 316: ......
Page 318: ......
Page 319: ......