SEP 2005
MODEL 1000
DESCRIPTION
1-12
In the quiescent condition (prior to injecting a sample), both legs of the bridge are exposed to pure
carrier gas. In this condition, the bridge is balanced and the bridge output is electrically nulled. (The
bridge can be balanced by the fine and coarse adjustment potentiometers located on the preamplifier
circuit board.)
The analysis begins when a fixed volume of sample is injected into the column by operation of the
sample valve. The sample is moved through the column by the continuous flow of carrier gas. As
successive components elute from the column, the temperature of the measurement element changes.
The temperature change unbalances the bridge and produces an electrical output proportional to the
component concentration. The differential signal developed between the two thermistors is amplified
by the preamplifier.
Figure 1-3 illustrates the change in detector electrical output during elution of a component.
Figure 1-3. Detector output during component elution
1.
Detector bridge balanced.
2.
First component begins to elute from column and to be sensed by the measurement
thermistor.
3.
Peak concentration of first component.
4.
Second component begins to elute from column and to be sensed by the measurement
thermistor.
5.
Peak concentration of second component.
In addition to amplifying the differential signal developed between the detector's two thermistors,
the preamplifier also supplies drive current to the detector bridge. The preamplifier also supplies
drive current to the detector bridge. The voltage signal is converted to a 4 to 20-milliamp (mA)
current loop for transmission to the GC Controller. The signal is proportional to the concentration
of a component detected in the gas sample. The preamplifier provides four different gain channels
as well as compensation for baseline drift. The signals from the preamplifier are sent to the GC
Controller for computation, recording on a printer, or viewing on a PC monitor or LCD.
Summary of Contents for Daniel Danalyzer 1000
Page 2: ......
Page 18: ...SEP 2005 MODEL 1000 DESCRIPTION 1 6 Figure 1 1 GC System Functional Block Diagram ...
Page 38: ...SEP 2005 MODEL 1000 EQUIPMENT DESCRIPTION 2 6 Figure 2 1 Analyzer Components ...
Page 146: ...SEP 2005 MODEL 1000 INSTALLATION AND SETUP 3 92 Figure 3 37 Analyzer Components ...
Page 188: ...SEP 2005 MODEL 1000 MAINTENANCE 5 20 Figure 5 2 Analyzer Components ...
Page 222: ...SEP 2005 MODEL 1000 MAINTENANCE 5 54 This page intentionally left blank ...
Page 228: ...SEP 2005 MODEL 1000 LIST OF APPENDICES A ii This page intentionally left blank ...
Page 260: ...SEP 2005 MODEL 1000 SETTING SOLENOID PURGE FLOWS E 4 This page intentionally left blank ...
Page 282: ...SEP 2005 MODEL 1000 ANALYZER DRAWINGS Addendum 1 2 This page intentionally left blank ...
Page 283: ......
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ......
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......
Page 299: ......
Page 300: ......
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......
Page 311: ......
Page 312: ......
Page 313: ......
Page 314: ......
Page 316: ......
Page 318: ......
Page 319: ......